OXY5500 Gas Analyzer Operating Instructions
Endress+Hauser 61
Bit Description
0 Interval on/off (delete to turn off, set to turn on)
1
Status-Bit: Set when a measurement is currently performed. will be deleted after completing the measurement.
Setting this bit does not trigger any action.
2 Start measurement (single scan or continuous)
3 - 31 Reserved
Table 16. Measurement control register bit definition
6.1.8 Compensation temperature
This value is used for compensation of the oxygen calculation.
Start Register Number of Registers Reg3 / Reg4 Write Access
2411 2 Float: Temperature value in °C Yes
Table 17. Definition of register 2411
6.1.9 Measurement interval
The oxygen measurement interval can be set between 1 and 359999. Setting the interval to “0” will lead to Modbus
error response with code 3.
The measurement values may be read any time but are updated only with the interval set in these registers. Hence,
polling the measurement values in a higher rate than the measurement interval should be omitted as it leads to
unnecessary traffic on the bus.
Start Register Number of Registers Reg3/Reg4 Write Access
3499 2 Int32: Interval value in seconds Yes
Table 18. Definition of register 3499
6.1.10 Device ID RS-485, RS-232 and Ethernet
Sets the device ID used in the Modbus RTU communication. If a value over 32 is set, the device will reset its ID to 1,
which may lead to communication errors. If no ID is set, or the ID is not known, set the ID via broadcast (ID=0).
Start Register Number of Registers Reg3/Reg4 Write Access
4095 2 Int32: Device ID of RS-485. Minimum 1, Maximum 32 Yes
Table 19. Definition of register 4095
Start Register Number of Registers Reg3/Reg4 Write Access
4109 2 Int32: Device ID of RS-232. Minimum 1, Maximum 32. Yes
Table 20. Definition of register 4109
Start Register Number of Registers Reg3/Reg4 Write Access
5695 2 Int32: Device ID of Ethernet. Minimum 1, Maximum 32. Yes
Table 21. Definition of register 5695
6.1.11 Measurement values
Start Register
# of
Registers
Reg1/
Reg2
Reg3/
Reg4
Reg5/
Reg6
Reg7/
Reg8
Reg9/
Reg10
Reg11/
Reg12
Reg13/
Reg14
Write
Access
4895 14
Float:
Pressure in
mbar
Float:
Reference
Amplitude
in mV
Float:
Oxygen
Amplitude
in mV
Float:
Oxygen
Phase shift
in degrees
Float:
Temperatu
re in °C
Float: Oxygen
Value
defined in
register 2089
Int32: Error
Register.
No
Table 22. Definition of register 4895