OXY5500 Gas Analyzer Operating Instructions
Endress+Hauser 63
Output Unit Sensor/Condition
Oxygen ppm gas OP-9
Temperature °C Always
Pressure Mbar Always
Table 27. Oxygen units for various output, sensor, and measurement mode configuration
Endress+Hauser recommends deactivating the current measurement before changing any settings. The device
will hold its last analog output value until the next measurement.
6.1.16 Ethernet IP
Start Register # of Registers Reg1/Reg2 Reg3/Reg4 Reg5/Reg6 Reg7/Reg8 Write Access
5677 8
Int32: Ethernet
IP Octet 1
Int32: Ethernet
IP Octet 2
Int32: Ethernet
IP Octet 3
Int32: Ethernet
IP Octet 4
Yes
Table 28. Definition of register 5677
Example:
Writing the following bytes:
0x 01 10 16 2D 00 08 10 00 C0 00 00 00 A8 00 00 00 01 00 00 00 0A 00 00 1F B1 will result in the IP 192.168.1.10
In detail:
0x 01 Slave address (int32 “01”)
0x 10 Function Code
0x 16 2D Starting Address (5677 in int32 representation)
0x 00 08 Quantity of registers
0x 00 C0 00 00 Octet 1 (int32 192)
0x 00 A8 00 00 Octet 2 (int32 168)
0x 00 01 00 00 Octet 3 (int32 1)
0x 00 0A 00 00 Octet 4 (int32 10)
0x 1F B1 CRC16
6.1.17 Ethernet subnet mask
Start Register # of Registers Reg1/Reg2 Reg3/Reg4 Reg5/Reg6 Reg7/Reg8 Write Access
5685 8
Int32: Ethernet
Subnet Mask
Octet 1
Int32: Ethernet
Subnet Mask
Octet 2
Int32: Ethernet
Subnet Mask
Octet 3
Int32: Ethernet
Subnet Mask
Octet 4
Yes
Table 29. Definition of register 5685