Operating Instructions OXY5500 Gas Analyzer
80 Endress+Hauser
8. Remove the conduit clamp screw with a Philips screwdriver. Refer to Figure 87.
Figure 87. Remove conduit clamp
9. Turn the conduit bracket parallel to the panel and carefully disengage the probe from the Tee union (panel side).
Refer to Figure 88.
Take care not to disturb the temperature probe while removing the oxygen probe conduit.
Figure 88. Remove probe from tee union (panel side)
10. Extend the probe conduit away from the panel and remove the fittings from the probe tip (panel side). Refer to
Figure 89.
Figure 89. Fittings on oxygen probe (panel side)
Take care to safely set aside the tube nut, cable gland cap, and plastic ferrules for use with the replacement probe.
Plastic
Ferrule
Tube
Nut
Cable
Gland
Cable
Gland Cap