Commissioning Proline Promass I 300 PROFINET
110 Endress+Hauser
Configuring the switch output
Navigation
"Setup" menu → Pulse/frequency/switch output
‣
Pulse/frequency/switch output
1 to n
Operating mode
→ 110
Terminal number
→ 110
Signal mode
→ 110
Switch output function
→ 111
Assign diagnostic behavior
→ 111
Assign limit
→ 111
Assign flow direction check
→ 111
Assign status
→ 112
Switch-on value
→ 112
Switch-off value
→ 112
Switch-on delay
→ 112
Switch-off delay
→ 112
Failure mode
→ 112
Invert output signal
→ 112
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Operating mode – Define the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
–
Terminal number – Shows the terminal numbers
used by the PFS output
module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Signal mode – Select the signal mode for the
PFS output.
• Passive
• Active
• Passive NAMUR
–