Proline Promass I 300 PROFINET Commissioning
Endress+Hauser 129
Parameter Prerequisite Description Selection / User
entry
Factory setting
Index
inhomogeneous
medium
Application
specific output 0
*
Index suspended
bubbles
*
0% bargraph value 1 A local display is provided. Enter 0% value for bar graph
display.
Signed floating-point
number
Country-specific:
0 kg/h
0 lb/min
100% bargraph value 1 A local display is provided. Enter 100% value for bar
graph display.
Signed floating-point
number
Depends on country
and nominal
diameter
Decimal places 1 A measured value is specified
in the Value 1 display
parameter.
Select the number of decimal
places for the display value.
x
x.x
x.xx
x.xxx
x.xxxx
Value 2 display A local display is provided. Select the measured value that
is shown on the local display.
For the picklist, see
the Value 2 display
parameter
(→  118)
Decimal places 2 A measured value is specified
in the Value 2 display
parameter.
Select the number of decimal
places for the display value.
x
x.x
x.xx
x.xxx
x.xxxx
Value 3 display A local display is provided. Select the measured value that
is shown on the local display.
For the picklist, see
the Value 2 display
parameter
(→  118)
0% bargraph value 3 A selection was made in the
Value 3 display parameter.
Enter 0% value for bar graph
display.
Signed floating-point
number
Country-specific:
0 kg/h
0 lb/min
100% bargraph value 3 A selection was made in the
Value 3 display parameter.
Enter 100% value for bar
graph display.
Signed floating-point
number
Decimal places 3 A measured value is specified
in the Value 3 display
parameter.
Select the number of decimal
places for the display value.
x
x.x
x.xx
x.xxx
x.xxxx
Value 4 display A local display is provided. Select the measured value that
is shown on the local display.
For the picklist, see
the Value 2 display
parameter
(→  118)
Decimal places 4 A measured value is specified
in the Value 4 display
parameter.
Select the number of decimal
places for the display value.
x
x.x
x.xx
x.xxx
x.xxxx