Proline Promass I 300 PROFINET Commissioning
Endress+Hauser 137
Parameter overview with brief description
Parameter Prerequisite Description Selection / User entry /
User interface
Assign simulation process variable – Select a process variable for the
simulation process that is activated.
• Off
• Mass flow
• Volume flow
• Corrected volume flow
*
• Target mass flow
*
• Carrier mass flow
*
• Target volume flow
*
• Carrier volume flow
*
• Target corrected volume
flow
*
• Carrier corrected volume
flow
*
• Density
• Reference density
*
• Temperature
• Dynamic viscosity
*
• Kinematic viscosity
*
• Temp. compensated
dynamic viscosity
*
• Temp. compensated
kinematic viscosity
*
• Concentration
*
Process variable value A process variable is selected in the
Assign simulation process variable
parameter (→ 137).
Enter the simulation value for the
selected process variable.
Depends on the process
variable selected
Status input simulation 1 to n – Switch simulation of the status input on
and off.
• Off
• On
Input signal level 1 to n In the Status input simulation
parameter, the On option is selected.
Select the signal level for the simulation
of the status input.
• High
• Low
Current input 1 to n simulation – Switch simulation of the current input
on and off.
• Off
• On
Value current input 1 to n In the Current input 1 to n simulation
parameter, the On option is selected.
Enter the current value for simulation. 0 to 22.5 mA
Current output 1 to n simulation – Switch the simulation of the current
output on and off.
• Off
• On
Value current output 1 to n In the Current output 1 to n
simulation parameter, the On option is
selected.
Enter the current value for simulation. 3.59 to 22.5 mA
Frequency output simulation 1 to n In the Operating mode parameter, the
Frequency option is selected.
Switch the simulation of the frequency
output on and off.
• Off
• On
Frequency value 1 to n In the Frequency output simulation
1 to n parameter, the On option is
selected.
Enter the frequency value for the
simulation.
0.0 to 12 500.0 Hz
Pulse output simulation 1 to n In the Operating mode parameter, the
Pulse option is selected.
Set and switch off the pulse output
simulation.
For Fixed value option: Pulse
width parameter (→ 106)
defines the pulse width of the
pulses output.
• Off
• Fixed value
• Down-counting value
Pulse value 1 to n In the Pulse output simulation 1 to n
parameter, the Down-counting value
option is selected.
Enter the number of pulses for
simulation.
0 to 65 535
Switch output simulation 1 to n In the Operating mode parameter, the
Switch option is selected.
Switch the simulation of the switch
output on and off.
• Off
• On
Switch status 1 to n – Select the status of the status output for
the simulation.
• Open
• Closed