Proline Prosonic Flow 91 Wiring
Endress+Hauser 23
a0005875
Fig. 23: Connecting nameplate for sensor connecting cables
4.1.2 Cable specifications
Sensor cable
Use the ready-to-use cables supplied by Endress+Hauser with each sensor pair.
The following cable lengths are available:
5 m, 10 m, 15 m , 30 m, 60 m
16 ft, 33 ft, 49 ft, 98 ft, 197 ft
Cable material: PVC
Operating temperature: –20 to +70 °C (–4 to +158 °F)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with
EN 61010-1, the EMC requirements of IEC/EN 61326 "Emission to class A requirements", and
NAMUR Recommendation NE 21.