Proline Prosonic Flow 91 Description of device functions
Endress+Hauser 101
11.19 Group SIMULATION SYSTEM
Functional description SIMULATION SYSTEM
SIMULATION
FAILSAFE MODE
Use this function to set all outputs and the totalizer to their defined failsafe modes, in order
to check whether they respond correctly.
During this time, the words "SIMULATION FAILSAFE MODE" appear on the display.
Options:
ON
OFF
Factory setting:
OFF
SIMULATION
MEASURAND
Use this function to set all outputs and the totalizer to their defined flow-response modes,
in order to check whether they respond correctly.
During this time, the words "SIMULATION MEASURAND" appear on the display.
Options:
OFF
VOLUME FLOW
Factory setting:
OFF
!
Note!
The measuring device cannot be used for measuring while this simulation is in progress.
The setting is not saved if the power supply fails.
VALUE SIMULATION
MEASURAND
!
Note!
This function is not available unless the SIMULATION MEASURAND function is active
(= VOLUME FLOW).
Use this function to define a freely selectable value (e.g. 12 m
3
/s).
This is used to test downstream devices and the measuring device itself.
User input:
5-digit floating-point number [unit], with sign
Factory setting:
0 [unit]
"
Caution!
The setting is not saved if the power supply fails.
!
Note!
The appropriate unit is taken from the SYSTEM UNITS group.