Commissioning Proline Prosonic Flow 91
42 Endress+Hauser
6.3.3 Data backup/transmission
Using the T-DAT SAVE/LOAD function ( ä 74), you can transfer data (device parameters and
settings) between the T-DAT (exchangeable memory) and the EEPROM (device storage unit).
This is required in the following instances:
Creating a backup: current data are transferred from an EEPROM to the T-DAT.
Replacing a transmitter: current data are copied from an EEPROM to the T-DAT and then
transferred to the EEPROM of the new transmitter.
Duplicating data: current data are copied from an EEPROM to the T-DAT and then transferred to
EEPROMs of identical measuring points.
!
Hinweis!
For information on installing and removing the T-DAT ä 55
a0001221-en
Fig. 31: Data backup/transmission with T-DAT SAVE/LOAD function
Information on the LOAD and SAVE options available:
LOAD: Data are transferred from the T-DAT to the EEPROM.
!
Hinweis!
Any settings already saved on the EEPROM are deleted.
This option is only available, if the T-DAT contains valid data.
This option can only be executed if the software version of the T-DAT is the same or newer than
that of the EEPROM. Otherwise, the error message "TRANSM. SW-DAT" appears after restarting
and the LOAD function is then no longer available.
SAVE:
Data are transferred from the EEPROM to the T-DAT
F
Esc
E
+
-
XXX.XXX.XX
F
F
F
F
F
F
P
P
P
P
N
O
T-DAT
SAVE/LOAD
Quick Setup
HOME
POSITION
LOAD
YES
NO
CANCELSAVE
YES
NO
Restart of the
measuring device
Input is
saved

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proline Prosonic Flow 91 and is the answer not in the manual?

Endress+Hauser Proline Prosonic Flow 91 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Prosonic Flow 91
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals