Proline Prosonic Flow 91 Description of device functions
Endress+Hauser 73
11.5 Group OPERATION
Functional description OPERATION
LANGUAGE Use this function to select the language for all texts, parameters and messages shown on
the local display.
Options:
ENGLISH
DEUTSCH
FRANCAIS
ESPANOL
ITALIANO
Factory setting:
Depends on country, see factory setting ä 103 ff.
!
Note!
If you press the OS keys simultaneously at startup, the language defaults to "ENGLISH".
ACCESS CODE All data of the measuring system are protected against inadvertent change.
Programming is disabled and the settings cannot be changed until a code is entered in this
function. If you press the P keys in any function, the measuring system automatically goes
to this function and the prompt to enter the code appears on the display (when
programming is disabled).
You can activate programming by entering your private code (factory setting = 91, see
also the subsequent DEFINE PRIVATE CODE function)
User input:
Max. 4-digit number: 0 to 9999
!
Note!
The programming levels are disabled if you do not press a key within 60 seconds
following automatic return to the HOME position.
You can also disable programming in this function by entering any number (other than
the defined private code).
The Endress+Hauser service organization can be of assistance if you mislay your
personal code.
DEF.PRIVATE CODE Use this function to enter a personal code to enable programming.
User input:
0 to 9999 (max. 4-digit number)
Factory setting:
91
!
Note!
This function only appears if the private code was entered in the ACCESS CODE
function.
Programming is always enabled with the code "0".
Programming has to be enabled before this code can be changed.
When programming is disabled this function is not available, thus preventing others
from accessing your personal code.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Proline Prosonic Flow 91 and is the answer not in the manual?

Endress+Hauser Proline Prosonic Flow 91 Specifications

General IconGeneral
BrandEndress+Hauser
ModelProline Prosonic Flow 91
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals