Proline Prosonic Flow P 500
Endress+Hauser 13
B
A
C
D D
C
C
D D
C
A0041975
 16 Two-path measurement: examples of the horizontal arrangement of the sensor sets at a measuring point
A Installation of the sensor sets for measurement via 1 traverse
B Installation of the sensor sets for measurement via 2 traverses
C For horizontal orientation: non-recommended installation range (60°)
D For horizontal orientation: recommended installation range max. 120°
If extending the measuring point from single-path measurement to two-path measurement, a
sensor of the same design must be selected.
Single-path measurement (alternative configuration)
In the case of single-path measurement, the flow is measured at the measuring point without the
option of compensation.
For this, it is necessary to comply strictly with the specified inlet and outlet runs after the
disturbance points (e.g. elbows, extensions, reductions) in the measuring pipe.
To ensure the best possible measurement performance and measurement accuracy, the
standard configuration with two sensor sets
4)
with FlowDC is recommended.
4) Order code for "Mounting type", option A2 "Clamp-on, 2-channel, 2-sensor sets"