Proline Prosonic Flow P 500
44 Endress+Hauser
Notes on mounting
Mounting the high-temperature sensors CH-050/CH-100
For detailed information on mounting the CH-050/CH-100 high-temperature sensors (order
code for "Sensor version", options AG, AH), see Special Documentation on "High-temperature
applications → 87".
Sensor configuration and settings
DN 15 to 65 (½ to 2½") DN 50 to 4000 (2 to 160")
Strapping band Strapping band Welded bolt
2 traverses
[mm (in)]
1 traverse
[mm (in)]
2 traverses
[mm (in)]
1 traverse
[mm (in)]
2 traverses
[mm (in)]
Sensor distance
1)
Sensor distance
1)
Sensor distance
1)
Sensor distance
1)
Sensor distance
1)
– Wire length → 52 Measuring rail
1) 2)
Wire length Measuring rail
1) 2)
1) Depends on the conditions at the measuring point (e.g. measuring pipe, medium). The dimension can be
determined via FieldCare or Applicator. See also Result sensor distance / measuring aid parameter in
Measuring point submenu
2) Up to DN 600 (24")
Determining the mounting positions of the sensor
Sensor holder with U-shaped screws )
Can be used for
• Measuring devices with measuring range DN 15 to 65 (½ to 2½")
• Mounting on pipes DN 15 to 32 (½ to 1¼")
Procedure:
1. Disconnect the sensor from the sensor holder.
2. Position the sensor holder on the measuring pipe.
3. Insert the U-shaped screws through the sensor holder and lightly grease the threads.
4. Screw the nuts onto the U-shaped screws.
5. Position the sensor holder exactly and tighten the nuts evenly.
A0043369
35 Holder with U-shaped screws
1 Sensor holder