Proline Prosonic Flow P 500
Endress+Hauser 53
Fastening the sensors:
1. Fit the sensor holders over the individual mounting bolts and tighten securely with the locking
nut.
13 mm
A0043381
 48 Mounting the sensor holders
2. Stick the coupling pad under the sensor →  87. Alternatively, coat the contact surfaces of
the sensor evenly with coupling gel (approx. 1 mm (0.04 in)). In doing so, start from the
groove through the center to the opposite edge.
A0043382
 49 Coating the contact surfaces of the sensor with coupling gel (if there is no coupling pad)
3. Insert the sensor into the sensor holder.
4. Fit the sensor cover on the sensor holder and turn until the sensor cover engages with a click
and the arrows (▴ / ▾ "close") are pointing towards one another.