Proline Prosonic Flow P 500
54 Endress+Hauser
5. Insert the sensor cable into each individual sensor until the end stop.
A0043383
 50 Mounting the sensors and connecting the sensor cables
This completes the mounting procedure. The sensors can now be connected to the
transmitter via the sensor cables and the error message can be checked in the sensor
check function.
To ensure good acoustic contact, the visible measuring pipe surface must be clean and free
from flaking paint and/or rust.
If the sensor is removed from the measuring pipe, it must be cleaned and new coupling gel
applied (if there is no coupling pad).
On rough measuring pipe surfaces, the gaps in the rough surface must be filled with
sufficient amounts of coupling gel if use of the coupling pad does not suffice (installation
quality check).
Installation for measurement via 2 traverses
Requirements
The installation distance is known. →  44
Strapping bands are pre-assembled
Material
The following material is required for mounting:
Two strapping bands incl. mounting bolts and centering plates where necessary (already pre-
assembled →  46, →  47)
A mounting rail to position the strapping bands:
Short rail up to DN 200 (8")
Long rail up to DN 600 (24")
No rail > DN 600 (24"), as distance measured by sensor distance between the mounting bolts
Two mounting rail holders
Two sensor holders
Coupling medium (coupling pad or coupling gel) for an acoustic connection between the sensor
and pipe
Two sensors incl. connecting cables
Open-ended wrench (13 mm)
Screw driver