3 Installation Proline Prowirl 72 FOUNDATION Fieldbus
16 Endress + Hauser
3.3 Installation
3.3.1 Mounting sensor
"
Caution!
Please note the following prior to mounting:
Prior to installing the measuring device in the piping, remove all traces of transport packaging and
any protective covers from the sensor.
Make sure that the internal diameters of seals are the same as, or greater than, those of the
measuring pipe and piping. Seals projecting into the flow current have a negative effect on the
vortex formation after the bluff body and cause inaccurate measurement. The gaskets provided
by Endress+Hauser for the wafer version have therefore an inner diameter with a bigger inner
diameter than the piping.
Ensure that the arrow on the measuring pipe matches the direction of flow in the piping.
Lengths:
Prowirl W (wafer version): 65 mm (2.56 inch)
Prowirl F (flanged version) See Technical Information TI070D/06/en.
Mounting Prowirl W
The centering rings supplied are used to mount and center the wafer-style devices.
A mounting kit consisting of tie rods, seals, nuts and washers can be ordered separately.
A0001888
Fig. 11: Mounting the wafer version
1Nut
2Washer
3Tie rod
4 Centering ring (is supplied with the device)
5Seal
1
2
3
4
5