3 Installation Proline Prowirl 72 FOUNDATION Fieldbus
12 Endress + Hauser
3.2.3 Orientation
The device can be installed basically in any orientation. Please consider the following, however
(Fig. 6):
In the case of liquids, there should be upward flow in vertical pipes to avoid partial pipe filling (see
orientation A).
The arrow on the meter body has to point in flow direction.
In order to make sure that the maximum ambient temperature (see Page 67) is not exceeded, we
recommend the following orientations:
In the case of hot fluids (e.g. steam or fluid temperature 200 °C/ 392 °F), select orientation
C or D.
Orientations B and D are recommended for very cold fluids (e.g. liquid nitrogen).
"
Caution!
If fluid temperature is 200 °C (392 °F), orientation B is not permitted for the wafer version
(Prowirl 72 W) with a nominal diameter of DN 100 (4") and DN 150 (6").
In case of vertical orientation and downward flowing liquid, the piping has always to be
completely filled.
A0001869
Fig. 6: Possible orientations of the device
High fluid temperature:
horizontal piping: orientation C or D
vertical piping: orientation A
Low fluid temperature:
horizontal piping: orientation B or D
vertical piping: orientation A
B
D
A
C