3.3.4 Mounting transmitter (remote)
The transmitter can be mounted in the following ways:
Wall mounting
Pipe mounting (with separate mounting kit, accessories Page 44)
The transmitter and the sensor must be mounted separate in the following circumstances:
poor accessibility,
lack of space,
extreme ambient temperatures.
"
Caution!
If the device is mounted to warm piping, make certain that the housing temperature does not
exceed the max. permissible temperature value.
Standard: –40 to +80 °C (–40 to +176 °F)
EEx d/XP version: –40 to +60 °C (–40 to +140 °F)
ATEX II 1/2 GD version/dust ignition-proof: –20 to +55 °C (–4 to +131 °F)
Mount the transmitter as illustrated in the diagram.
A0003801
Fig. 14: Mounting the transmitter (remote version)
A Direct wall mounting
BPipe mounting
* Dimensions for version without local display
ANSCHLUSSKLEMMEN - FIELD TERMINALS
ANSCHLUSSKLEMMEN - FIELD TERMINALS
A
B
220 / *214
215/ *209
(8.66 / *8.43)
(8.46 / *8.23)
mm (inch)
Ø 20…70
(Ø 0.79…2.75)