Operating the Analyzer
Operator’s Manual 2
23
Table 2-3 Assignable Alarm functionality (Continued)
Bit Value Functionality
4 0000010 Laser zero is too low.
5 0000020 Laser zero is too high.
6 0000040 Laser current is too low.
7 0000080 Laser current is too high.
8 0000100 Pressure is too low.
9 0000200 Pressure is too high.
10 0000400 Temperature is too low.
11 0000800 Temperature is too high.
12 0001000 Concentration is too low.
13 0002000 Concentration is too high.
14 0004000 Peak tracking restart exceeds limit.
15 0008000 Fitting restart exceeds limit.
16 0010000 Ramp adjust restart exceeds limit.
17 0020000 Not used.
18 0040000 Not used.
19 0080000 Flow Switch Alarm.
20 0100000 Validation Fail Alarm 1.
21 0200000 Validation Fail Alarm 2.
22 0400000 Not used.
23 0800000 Not used.
24 1000000 Delta DC restart exceeds limit.
25 2000000 Delta temperature restart exceeds limit.
26 4000000 Dry pressure restart exceeds limit.
27 8000000 Need to change scrubber.
28 10000000 R2 restart exceeds limit.
29 20000000 R3 restart exceeds limit.
30
40000000 Pressure delta restart exceeds limit.
31 80000000 Low purge rate alarm.

Table of Contents

Other manuals for Endress+Hauser SpectraSensors SS2100i-1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors SS2100i-1 and is the answer not in the manual?

Endress+Hauser SpectraSensors SS2100i-1 Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors SS2100i-1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals