Val Auto DumpSpectrm
The Val Auto DumpSpectrm parameter determines whether a Mode 6 dump
automatically occurs after each validation measurement. There are two
choices: 0 to turn the automatic data dump during validation off, and 1 to turn
it on.
Validation Allowance
The Validation Allowance parameter sets the tolerance (%) for validation
measurement when Val 1 Concentration or Val 2 Concentration is set to a
value greater than 0.
Validation Wait Time
The Validation Wait Time parameter sets the amount of time (in seconds)
for the validation gas to purge the system at the beginning of the validation.
Because validation gas may be introduced into the system at various distances
from the analyzer, adjustment of the Validation Wait Time parameter is
necessary to optimize the time the validation gas is allowed to purge through
the transport tubing before the analyzer makes a validation measurement.
Optimization of the Validation Wait Time parameter ensures an accurate
measurement of the validation gas while minimizing gas consumption.
<SET PARAMETER MODE>
Val Attempts
2
Enter a value
<SET PARAMETER MODE>
Val Auto DumpSpectrm
0
0:Disable 1:Enable
<SET PARAMETER MODE>
Validation Allowance
10.00000
% of val concentra