NS 5.14 Firmware
34 4900002262 rev. B 3-23-21
Concentration: Current measured analyte concentration (ppmv).
Wet Temp: Current temperature of the gas sample (degrees C)
when normal sample gas is flowing.
Wet Pressure: Current pressure of the gas sample (mb) when
normal sample gas is flowing.
Dry Temp: Current temperature (0.0 for non-differential units) of
the gas sample (degrees C) when scrubbed sample gas is flowing.
Dry Pressure: Current pressure (0.0 for non-differential units) of
the gas sample (mb) when scrubbed sample gas is flowing.
Fit Residue: Value ranging from 0 to 1 indicating how well the
measured spectrum fits the reference spectrum, where 1 represents
a perfect match.
Fit Ratio: Ratio of the measured spectrum versus the Reference 1
spectrum.
Dry DC: Signal level (laser intensity in counts, 0.0 for non-
differential units) at the high end of the current ramp during the dry
cycle.
Wet DC: Signal level (laser intensity in counts) at the high end of the
current ramp during the wet cycle.
Peak Index: Peak index of the measured spectrum.
Ref Index: Peak index used for reference.
Index Difference: Difference between measured and reference
peak indices, where a value other than 0 indicates Peak Tracking is
functioning.
Val Flg: Indicates current stream being measured (0 = Process,
1 = Val 1, 2 = Val 2).
Process Path Flg: Calculation path (reference spectra) used for the
concentration calculation (0 = Process Reference Spectra,
1 = Validation Reference Spectra).
Current Midpoint: Current midpoint that the analyzer is using,
including any peak tracking adjustments.
Fit Ratio 2: Ratio of the measured spectrum versus the Reference 2
spectrum, where a value of 0 indicates that the Reference 2 spectrum
was not used in the calculation of concentration.
Fit Ratio 3: Ratio of the measured spectrum versus the Reference 3
spectrum, where a value of 0 indicates that the Reference 3 spectrum
was not used in the calculation of concentration.
Fit Ratio 4: Ratio of the measured spectrum versus the Reference 4
spectrum, where a value of 0 indicates that the Reference 4 spectrum
was not used in the calculation of concentration.
Fit Ratio 5: Ratio of the measured spectrum versus the Reference 5
spectrum, where a value of 0 indicates that the Reference 5 spectrum
was not used in the calculation of concentration.