NS 5.14 Firmware
3–18 4900002262 rev. B 3-23-21
1.Write privilege requires User Level 1 (L1) password to be downloaded to the
Password register.
Table 3-1 Modbus register map (Continued)
Parameter
Daniel
Register
Gould
Register
Data
Type
Action Minimum Maximum
4-20 mA Alarm Action 3204 43204 Integer
R/W L1
1
03
Temperature Unit 3205 43205 Integer
R/W L1
1
01
Pressure Unit 3206 43206 Integer
R/W L1
1
03
Concentration Unit 3207 43207 Integer
R/W L1
1
08
Modbus Address 3208 43208 Integer
R/W L1
1
0250
Modbus Mode 3209 43209 Integer
R/W L1
1
02
AI Pressure Input 3210 43210 Integer
R/W L1
1
01
RATA 3211 43211 Integer
R/W L1
1
01
General Alarm DO 3213 43213 Integer
R/W L1
1
02
Baud Rate 3214 43214 Integer
R/W L1
1
04
Set Time - Year 3215 43215 Integer
R/W L1
1
2007 2143
Set Time - Month 3216 43216 Integer
R/W L1
1
112
Set Time - Day 3217 43217 Integer
R/W L1
1
131
Set Time - Hour 3218 43218 Integer
R/W L1
1
023
Set Time - Minute 3219 43219 Integer
R/W L1
1
059
Cancel Val Alarms 3220 43220 Integer
R/W L1
1
01
Daily Validation 3221 43221 Integer
R/W L1
1
01
Val Start Time 3222 43222 Integer
R/W L1
1
023
Val Interval 3223 43223 Integer
R/W L1
1
1400
Start Validation 3224 43224 Integer
R/W L1
1
01
Val Purge Period 3225 43225 Integer
R/W L1
1
0 4000
Val Duration 3226 43226 Integer
R/W L1
1
0 8000
Val Attempts 3227 43227 Integer
R/W L1
1
1 8000
4-20 mA Val Action 3228 43228 Integer
R/W L1
1
01
Val Auto DumpSpectrum 3230 43230 Integer
R/W L1
1
01
Update RATA 3231 43231 Integer
R/W L1
1
01
Conversion Type 3251 43251 Integer
R/W L1
1
01
Calculate Dew Point 3252 43252 Integer
R/W L1
1
02