NS 5.14 Firmware
330 4900002262 rev. B 3-23-21
Validation Allowance: Sets the tolerance (±%) for validation
measurements when Val 1 Concentration or Val 2 Concentration is set
to a value greater than 0.
Validation Wait Time: Sets the amount of time (in seconds) for the
validation gas to purge the system at the beginning of the validation.
Val Perm Constant Kp: Sets the system constant for the perm
tube.
Val Perm Rate Rp: Sets the calibrated permeation rate for the perm
tube.
Val Purge Period: Sets the number of seconds upon validation
initiation that a differential analyzer will purge the system with
validation gas before starting a dry cycle.
Val Start Time: Sets the hour of the day for the daily autovalidation
to begin.
Val Time: Time of last validation.
Wet Pressure (mb): Current measured (live) wet pressure reading
of the gas sample in millibar (mb).
Wet Temp (C): Current measured (live) temperature of the gas
sample in degrees Celsius.
Zero Val Tolerance: Sets the maximum acceptable reading (in
ppmv) when validating with zero gas (Val 1 Concentration = 0 or Val
2 Concentration = 0).
Zero Level: Signal level when the laser is turned off.

Table of Contents

Other manuals for Endress+Hauser SpectraSensors SS2100i-1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors SS2100i-1 and is the answer not in the manual?

Endress+Hauser SpectraSensors SS2100i-1 Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors SS2100i-1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals