Checks and Maintenance
42
Bench Top User’s Manual
7.1 Check and Maintenance Lists
Check List
For an explanation on each check item, See “Checks” on page 43. If any of the following checks result in
problems, contact the place of purchase or Espec North America.
Maintenance List
For an explanation on each maintenance item, See “Maintenance” on page 45.
*Not required on humidity models with solid state humidity sensor option
**Humidity models (BTL, BTX) only
***Not required on models with liquid nitrogen option BT-1LN
TABLE 7.1.
CHECK LIST
Check Item When to Check
Overheat Protector Trip Test
Before long test runs
Before unattended tests
Wet-Bulb Wick Check* Before each test
Wick Pan Water Level Check
Once every 3 months
After moving chamber. See “Checking
Water Level in the Wick Pan” on page 29.
TABLE 7.2. MAINTENANCE LIST
Maintenance Item When to Perform
Condenser Cleaning*** Once Monthly
Steam Generator Cleaning** Once Monthly
Test Area Cleaning Before Every Test
Take-Down Before Long Periods of Disuse