Bench Top User’s Manual
Appendix
9.2 Consumable Parts and Replacement Interval
The below parts must be replaced periodically. Promptly replace them at the specified interval if not
sooner. Maintenance service is also provided by ESPEC North America. For spare parts and servicing,
contact the place of purchase or Espec North America..
TABLE 9.5. CONSUMABLE PARTS AND REPLACEMENT INTERVAL
Part Replacement Interval Replacement Procedure
Door Gasket 3 Years Contact the place of purchase or ESPEC.
Light Bulb
1-2 Years depending on use See “Changing the Chamber Lamp Bulb”
on page 55.
Humidity Wick
2-4 weeks depending on water quality See “Wet Bulb Wick Check (Not neces-
sary with temperature-only chambers)” on
page 43