Checks and Maintenance
48
Bench Top User’s Manual
7.3.3 Test Area Cleaning
Dirt and foreign matter inside the test area can throw test results off. Clean the test area before every test.
Procedure:
1. Open the chamber door.
2. Wipe walls and parts clean with a soft cloth.
3. Close the door.
7.3.4 Take-Down Before Long Periods of Inactivity
Before inactivity of 3 days or more, dry the test area to prevent mildew or scale formation. Failure to do so
can affect testing and shorten equipment service-life.
Procedure:
1. Check that the main power cord is plugged in.
2. Set target temperature to 20°C and turn OFF humidity control.
3. Turn the Chamber Run switch ON.
4. Run the cabinet for about 60 minutes with the door closed.
5. Turn the Chamber Run switch OFF.
6. Open chamber door and leave ajar.
7. Unplug chamber power cord.
8. Drain steam generator.