3. Site requirements
3.1. Ambient Temperature Allowable range of operation: 5 to 35°C (41 to 95°F)
Range of assured performance: 18 to 27°C (65 to 80°F)
3.2. Power (Full Load Current)
Model
115 V 1 ph 60 Hz
3 wire grounding
BTL-433
16
3.3. Refrigeration Cooling
3.3.1 Air Cooled (Integral)
** Noise level measured 1 meter from front and at mid-point of door
3.4. Gravity Drain 1/2” Hose barb
3.5 Humidity Water
3.5.1. Connection 3/8” Hose barb
3.5.2. Flow
4
Peak 1 L/min (30-50 psig)
3.5.3. Water Quality 0.2 µS/cm to 10 µS/cm.
Maximum 2 mg/L free chlorine & filtered to 5 micron or less.
4
Initial fill of an empty steam generator. See detailed specification document for actual humidity usage rate.
Model Heat Rejection to Ambient Noise Level**
BTL-433 3300 BTU/Hr 65 dBA

Table of Contents