E8_Aus_rev07 – 2021-02-02
Mechanical safety devices
Door safety micro-switch: it ensures correct closure
of the door
Message indicating wrong door position
Door-lock micro-switch: it detects the correct
position of the locking system
Indication that the door is not locked
Door lock: electro-mechanical device that prevents
the door from being opened accidentally
It prevents the door from being opened while the unit
is in operation.
Attempting to open the door with the
door safety device applied may
seriously damage the closing system
Extractor tool. Used to avoid touching the inner parts
of the unit.
Prevents burns while removing the trays containing
the sterilized instruments
Control devices
Pressure levelling: restores the system to its normal
pressure values, in the event of manual stops or
alarms and/or warnings during the cycle.
Automatic pressure re-balancing inside the
sterilization chamber.
System for evaluating process parameters,
managed entirely by the microprocessor
In the event of faults during the cycle, the program in
progress is stopped immediately and alarms are
generated
Constant monitoring of the device: the steriliser
components are constantly monitored during
operation.
Generation of alarm messages and/or warnings in
the event of faults
It is forbidden to remove, modify, tamper with or neutralize the safety devices in any way.
Euronda S.p.A. declines all liability for accidents to people or damage or malfunctions to
the unit if the above instruction is not observed.
Periodically check the safety systems (see chap. 10 “Maintenance”).