E8_Aus_rev07 2021-02-02
33
Door release
ATTENTION: A safety pin automatically locks the door when the cycle starts. The pin only returns
to its seat at the end of the cycle. Attempting to open the door with the door safety device
applied may seriously damage the closing system. Always wait for the end of cycle signal
on the LCD display before opening the door.
In the event of an alarm, the door can only be opened after giving consent by touching the middle
button (see 7.6).
CAUTION: LOAD NOT STERILE, HANDLE WITH CARE.
Do not release the door safety device manually
7.2.2 Information on process parameters
Additional information on the parameters of the currently running cycle can be obtained by pressing "i" on the
Cycle in progress screen.
The next screen shows the type of cycle currently running, and the number of cycles completed. Below is the
instantaneous reading of the probes on the steriliser and the name of the current phase (at the top). Below
that, is the icon with the left arrow, to return to the previous screen.
* If Aquafilter is set to ON, the conductivity measure appears.
7.2.3 How to release the Personal cycles
Note: In order to receive the password and release the Personal Light and Light&Stock cycles, you
need to register your E8 on www.myeuronda.com.
CAUTION: use these sterilisation programs only for solid loads and not for hollow and double-
packed ones; read the description of the programs in appendix 5 with care