- Pursuant to Article 33 of Regulation 1907/2006 (REACH), it should be noted that the unit contains
Decamethylcyclopentasiloxane, CAS 541-02-6. This substance can not come into contact with the user or
to the sterilized tools as it is contained inside the heating element, therefore, no associated hazards
and/or precautions for use.
- Pursuant to Article 33 of Regulation 1907/2006 (REACH), it should be noted that the unit contains
Dodecamethylcyclohexasiloxane, CAS 540-97-6. This substance can not come into contact with the user
or to the sterilized tools as it is contained inside the heating element, therefore, no associated hazards
and/or precautions for use.
- Pursuant to Article 33 of Regulation 1907/2006 (REACH), it should be noted that the unit contains
Octamethylcyclotetrasiloxane, CAS 556-67-2. This substance can not come into contact with the user or
to the sterilized tools as it is contained inside the heating element, therefore, no associated hazards
and/or precautions for use.
3.2 INTENDED USE
E8 water steam steriliser: unit designed and developed for the sterilisation of instruments used in
medical, dental, veterinary and podology surgeries, and tattoo shops, that can be sterilised by steam
between 121°C and 134°C.
The unit is designed for use by professional, qualified persons only. The unit must only be used for the
purpose it was designed for.
The manufacturer cannot be held responsible for any breakage, damage or malfunctioning
of the unit if the machine has not been used correctly, been used inappropriately or not
adequately maintained.
3.3 SAFETY DEVICES
Electrical safety
Double-pole thermal protection circuit breaker for
protection against short-circuit.
Disconnects main electrical power supply
Protection of the electronic card against short-circuit:
both the transformer and the entire low-voltage
circuit are self-protected.
Disconnects one or more low-voltage circuits
Thermal protection
Thermal protection of the unit: if anomalous
overheating occurs, the unit turns off.
The unit cannot be switched on until the manually
resettable thermostat on the front panel is reset.
The vacuum pump is protected by an automatically
resettable thermostat.
Temporary cut-off to permit cooling
The vibration pump is protected by a thermal fuse.
Interruption of electrical circuit, replace the thermal
fuse.
Manually resettable thermostat protecting the steam
generator
Interruption of power input to the steam generator
until the thermostat on the generator is reset.
Safety valve, complying with the PED 2014/68/EC
standards, for protecting the unit from over-pressure
Discharge of steam and re-balancing of pressure to
safety values