E8_Aus_rev07 2021-02-02
66
APPENDIX 8
Quality of process water
With reference to standard EN 13060, the table below indicates the recommended limit values (maximum)
for contaminating agents, as well as the chemical-physical characteristics of the water used for condensate*
and inlet water.
* Condensate is produced by the steam that was formed by the empty chamber of the sterilizer.
Supply water
Condensate
Dry residue
<10 mg/l
<1 mg/l
Silicon oxide
≤1 mg/l
≤0.1 mg/l
Iron
≤0.2 mg/l
≤0.1 mg/l
Cadmium
≤0.005 mg/l
≤0.005 mg/l
Lead
≤0.05 mg/l
≤0.05 mg/l
Heavy metal residues
≤0.1 mg/l
≤0.1 mg/l
Chlorides
≤2 mg/l
≤0.1 mg/l
Phosphates
≤0.5 mg/l
≤0.1 mg/l
Conductivity at 20°C
≤15 µS/cm
≤3 µS/cm
pH
5-7
5-7
Appearance
colourless, clean,
sediment-free
colourless, clean,
sediment-free
Hardness
≤0.02 mmol/l
≤0.02 mmol/l
NOTE: The use of water for generating steam containing contaminants at higher levels than those
shown in this table may considerably shorten the working life of a sterilizer and may invalidate the
maker’s guarantee.