110
Eversense E3 CGM User Guide
9
Alert Descriptions and Actions (continued)
Alerts
App DisplayDescription
Out of Range Low Glucose*
Appears when your glucose reading is lower than
40 mg/dL.
No glucose readings can be displayed (only LO
is displayed on the MY GLUCOSE screen).
Out of Range High Glucose*
Appears when your glucose value is higher than
400 mg/dL.
No glucose readings can be displayed (only HI
is displayed on the MY GLUCOSE screen).
Actions
Confirm your glucose value with a blood glucose
meter test before making a treatment decision.
Once the sensor glucose value is at or higher than
40 mg/dL, glucose readings will resume on the
display.
Confirm your glucose value with a blood glucose
meter test before making a treatment decision.
Once the sensor glucose value is at or lower than
400 mg/dL, glucose readings will resume on the
display.