EasyManua.ls Logo

eversense E3 - Calibration Phases

Default Icon
241 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
Eversense E3 CGM User Guide
6
6. Calibrating the System
This section describes the calibration procedure and schedule of your Eversense E3 CGM System.
To ensure best performance, routine calibration is required using fingerstick readings from a blood glucose meter.
Any commercially available meter may be used for calibration. Once your sensor has been inserted and linked to
your smart transmitter, the system begins a 24-hour Warm-Up Phase. No calibration is required during this phase.
There are three calibration phases:
Initialisation Phase – After the 24-hour Warm-Up Phase, you must complete 4 fingerstick calibration tests,
spaced 2 to 12 hours apart.
2 Daily Calibrations Phase – After the Initialisation Phase, you must complete a fingerstick calibration test every
12 hours for at least 21 days.
1 Daily Calibration Phase – The system will detect if 1 or 2 daily calibrations are needed and will alert you. In the
1 Daily Calibration Phase, you must complete a fingerstick calibration test every 24 hours.
Warning: DO NOT use alternative test sites such as your forearm when entering BG values for calibration.
After the first 21 days, the system will notify you if
1 or 2 daily calibrations are needed.
6

Table of Contents

Other manuals for eversense E3

Related product manuals