Eversense E3 CGM User Guide
6
Warning: In the 2 Daily Calibrations Phase, the Eversense E3 CGM System requires calibration every 12 hours
in order to provide accurate readings. In the 1 Daily Calibration Phase, the Eversense E3 CGM System requires
calibration every 24 hours. You should not use CGM readings to make treatment decisions unless you have
followed the instructions for daily calibration.
Routine calibration is critically important to ensuring the best performance of the Eversense E3 CGM System.
The following tips can help you improve your calibration measurements:
Tips for ensuring good calibration:
Calibrate at times when glucose is NOT changing rapidly (e.g., before meals, before dosing insulin).
Calibrate when you know you will not be removing the smart transmitter during the next 15 minutes.
Wash your hands with warm, soapy water and dry thoroughly before taking a blood glucose meter reading. It is
very important to have clean, dry hands when you test your blood glucose.
Always follow the blood glucose meter manufacturer’s instructions to get accurate blood glucose readings for
calibration.
Be sure the code on the test strip vial matches the code on your blood glucose meter (if coding is required).
Calibration will NOT be complete or results NOT accepted if:
Blood glucose meter reading is less than 40 mg/dL.
Blood glucose meter reading is greater than 400 mg/dL.
Blood glucose meter reading was taken more than 10 minutes before entering the result in the Eversense App.
Sensor glucose reading is significantly dierent than the blood glucose meter reading.
Your smart transmitter was removed or could not collect sensor glucose data during the 15 minutes after you
entered your calibration value.