80
Eversense E3 CGM User Guide
8
Setting Glucose Alert Levels
Your Eversense E3 CGM System will alert you when your glucose levels are outside the alert settings you choose.
When you have passed your low and high glucose alert levels, the smart transmitter vibrates, and the mobile app
gives an audible alert as well as displays a message on the screen.
Default setting Low: 65 mg/dL
High: 250 mg/dL
You can change these alert levels based on what you and your GP agree are
the right levels for you. Your Low Glucose Alert cannot be set above your Low
Glucose Target, and your High Glucose Alert cannot be set below your High
Glucose Target.
Allowable setting Low: 60-115 mg/dL
High: 125-350 mg/dL
On/O setting Low Glucose Alert setting is Always ON
High Glucose Alert setting can be turned ON and OFF. No High Glucose alerts
will display or vibe on the smart transmitter if this feature is turned o.
Notes
Audio notification and visual alerts on your mobile device and smart transmitter
on-body vibe alerts.
IMPORTANT:
The Low and High Glucose Alert levels you set are the same levels used to provide Predictive Alerts. See this
section, Setting Predictive Alerts.
For iOS 12 and above, and Android 6 and above, you can allow the Eversense Low Glucose and Out of Range Low
Glucose Alerts to override your phone sound settings. See Sound Settings for more information.