116
Eversense E3 CGM User Guide
9
Alert Descriptions and Actions (continued)
Alerts
App DisplayDescription
Sensor Suspend
Appears when the system’s internal checks detect a
need to restart the Initialisation Phase for additional
calibrations.
No glucose readings can be displayed until after the
second successful calibration.
Calibration Past Due
Appears when your system is past due for calibration
(16 hours in 2 Daily Calibrations Phase and 28 hours
in 1 Daily Calibration Phase).
No glucose readings can be displayed until calibration
is performed.
Actions
Re-initialisation of the system begins in 6hours.
Glucose values will be displayed a few minutes
after the second successful calibration during the
Initialisation Phase.
Perform a fingerstick calibration in order to resume
displaying glucose values.