19
205
Eversense E3 CGM User Guide
205
Eversense E3 CGM User Guide
*Glucose values between 40 and 400 mg/dL.
Table 1 – Accuracy to YSI in PROMISE*
Percent of CGM System Readings Within MARD
Total Number
of Paired
CGM and YSI
Values
Percent
15/15%
of
Reference
Percent
20/20%
of
Reference
Percent
30/30%
of
Reference
Percent
40/40%
of
Reference
12,034 87.3 93.9 98.6 99.6 8.5%