Electrical and Safety Standards (continued)
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Eversense E3 CGM Mobile System is intended for use in the electromagnetic environment specified in the next
table. The customer or the user of the System should ensure that it is used in such an environment.
Emissions
Test
Compliance
Electromagnetic Environment
Guidance
RF Emissions
CISPR 11
Group 1
The Eversense E3 CGM System uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B
The Eversense E3 CGM System is suitable for use in all
establishments including domestic and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Recommended Separation Distances Between Other Portable/Mobile RF Communications Equipment
and the Smartphone (Receiving Device)
Follow the smartphone (or other receiving device) manufacturer’s instructions for separation distances. The
customer or the user of the smartphone (or other receiving device) can help prevent electromagnetic interference
by maintaining a minimum distance between other portable/mobile RF communications equipment (transmitters)
and the smartphone of at least 30 cm (about 12 inches). Portable/mobile RF equipment include: baby monitors,
Bluetooth wireless headsets, wireless routers, microwave ovens, laptops with internal Wi-Fi adapters, GSM mobile
phones, RFID scanners and hand-held security metal detector often used by security screeners.