38
Eversense E3 CGM User Guide
4
Note: The sensor requires a 24-hour Warm-Up Phase to stabilise in your body
before glucose values will be collected by the smart transmitter. During the
Warm-Up Phase, you do not need to wear the smart transmitter. If you decide
to wear the smart transmitter over the sensor during this time, you will receive
a message on the app indicating the Warm-Up Phase is in progress. Once the
Warm-Up Phase is complete, turn ON the smart transmitter and place it over
the sensor with the Eversense adhesive patch. The system will prompt you to
calibrate using the app.
After the sensor is linked, you can remove the smart transmitter and put it in
the charging cradle for the next 24 hours. See the section, Using the Smart
Transmitter, to learn about using your system after the Warm-Up Phase.
IMPORTANT: After the 24-hour Warm-Up Phase, if your smart transmitter is not turned on and worn
over the sensor, the system cannot provide glucose readings after the Warm-Up Phase.
Caution: The sensor and smart transmitter should be linked the day of the
sensor insertion. Failure to link the sensor and smart transmitter could result in
a delay in receiving glucose readings.
4
Warning: The Eversense E3 CGM System will not provide readings during the 24-hour Warm-Up Phase
and until a second calibration is successful during the Initialisation Phase. During this time, you should
monitor your glucose using a home blood glucose monitor.
Tip: Your sensor may not be precisely perpendicular to the incision. If you find
it dicult to get a Good or Excellent signal in the Placement Guide, DO NOT apply
pressure. Do try slightly rotating the smart transmitter over the sensor. Wait about
1 second for the Placement Guide to refresh between each adjustment to the smart
transmitter’s position over the sensor.