44
04011210
EXCELL PRECISION CO., LTD
ZSME400000003
Command mode
1. Command Format A
Host Command
<CR>< LF>
Slave
Command
<CR>< LF>
MZ Zero CZ Zero compensation On/OFF
MT Tare CT Clear TARE value
MG Gross Weight MN Net weight
AT Accu. Current net weight and times plus 1
ST Deduct times of last accu. Value minus 1
DT Clear accu. Value and times
BB
Start batching (one time)
HB Load stop
BC Start batching (continuous)
BD Start unload
SC Transmit continuous SA Auto transmit
SM Manual transmit SO Command mode
% Stop continuous transmission and enter the command mode
Note Since BC continuous weighing is completed, it must zero re-set the
continuous weighing frequency by BB command.
2. Command Format B
Host Command
<CR>< LF>
Slave
Data
<CR>< LF>
RW Read current weight RT Read TARE
RG Read Gross Weight RN Read Net weight
RB Read current display of wt (simple) RH Read Gross (simple)
RI Read Net (simple)
RJ Read comparison situation + current display of weight (simple)
RK Read comparison situation + Gross (simple)
RL Read comparison situation + Net (simple)
RO Read comparison situation (2)
RF Read prior completed weight RA Read accu. Value (incl. times)
Note Prior command plus %
Read Weight Compared value: RS□□
FW
Read target item of unload value
S1 Read SP1
S2 Read SP2 S3 Read SP3
UD Read Under LO Read LO
ZB Read Zero Band HI Read HI
PR Reading Peak value OV Read Over
Ex:
Command : RSFWCR><LF Indicator reply : RSFW□□□□□□
Finish 6 bytes