Test Setup - Test Configurator, Timer, and System
136 FTB-700G/800 Series
EtherSAM - Burst
Parameters
Number of Burst Sequence is the number of times, 1 to 100 (default
is 2), the burst sequence will be repeated for the CBS and EBS tests.
Refill Delay Ratio (%) is the percentage of time required to refill the
CBS/EBS token buckets. The refill delay ratio is used for post-burst
delay and the remaining percentage is applied to the pre-burst delay.
The Refill Delay Ratio is configurable from 0 to 100 percent
(50 percent is the default as well as the standard’s minimum
recommended value).
Burst/IR Frame Ratio (%) is the percentage of frames transmitted at
CIR rate for the CBS test and the percentage of frames transmitted at
CIR+EIR rate for the EBS test. The Burst/IR Frame Ratio is
configurable from 10 to 90 percent (90 percent is the default as well as
the standard’s recommended percentage value).
Table
Note: Test time values are only displayed for enabled services.
Service No indicates the service number.
Service Name indicates the name of the service.
Direction, available with Dual Test Set, indicates results from local to
remote (L->R) and remote to local (R->L).
CBS Test Time (s) indicates the total time required to perform all burst
sequence iterations of the CBS test for this service.
EBS Test Time (s) indicates the total time required to perform all burst
sequence iterations of the EBS test for this service.
Total Burst Test Time (s) indicates the total time required to perform
all burst sequence iterations of both CBS and EBS tests for this service.