• Fix the extension cable and the external shield to the shield terminal.
• Connect the extension cable to the terminals of the junction box, see Fig. 7.7 and Tab. 7.5.
Important!
The external shield of the extension cable must not have electrical contact to the junction box. Therefore, the extension
cable has to remain completely insulated up to the shield terminal.
Tab. 7.5: Terminal assignment
terminal connection (extension cable)
TV white or marked cable (core)
TVS white or marked cable (internal shield)
TRS brown cable (internal shield)
TR brown cable (core)
shield terminal external shield
Fig. 7.7: Connection of the extension and transducer cable to the junction box
1 – extension cable
2 – external shield
3 – shield terminal
4 – cap nut
5 – compression part
6 – basic part
7 – connection of the extension cable
8 – connection of the transducer cable
cable gland
4 5 6
100 mm
20 mm
10 mm
1
2
3
7 8