EasyManua.ls Logo

Flexim FLUXUS F501 - Page 57

Default Icon
146 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Connection
FLUXUS F501 7.1 Transducers
57
UMFLUXUS_F501V1-3EN, 2020-06-30
7.1.5 Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced, the sensor module must also be replaced.
Disconnect the transmitter from the power supply.
Insert the transducer module into the socket, see Fig. 7.8.
Connect the transmitter to the power supply.
Enter all parameters of the program branch Parameter.
Start the measurement.
Notice!
The serial numbers of the transducer module and the transducer must be identical. A wrong or incorrectly connected
sensor module will lead to incorrect measured values or to a measurement failure.
Fig. 7.8: Connection of the transducer module
1 transducer module
AV
AVS
AGN
ARS
AR
1

Table of Contents

Related product manuals