9 Measurement
9.2 Measurement settings FLUXUS F501
2020-06-30, UMFLUXUS_F501V1-3EN
78
Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
Full-Scale Val. is the value assigned to the value Output MAX of the output range.
Press ENTER.
Function test
The function of the output can now be tested.
Connect an external measuring instrument to the terminals of the output.
Select yes to test the output.
Press ENTER.
Enter a test value for the selected physical quantity. If the external measuring instrument displays the entered value, the
output functions correctly.
Press ENTER.
Select yes to repeat the test.
Press ENTER.
Output Options\...\Full-Scale Val.
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA
Output Options\...\I1:Test output?
Output Options\...\I1:Test value
Output Options\...\I1:Test output?
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 m³/h
Test value = 150 m³/h (center of the measuring range, corresponds to 12 mA)
If the multimeter displays 12 mA, the current output functions correctly.