18 Settings
18.1 Dialogs and menus FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
158
18.1.6 Alarm state indication
• Select on to display the alarm state during the measurement.
• Press ENTER.
18.1.7 Units of measurement
It is possible to set the units of measurement for the length, temperature, pressure, density and kinematic viscosity and
sound speed.
• Select a unit of measurement for all quantities.
• Press ENTER after each selection.
18.1.8 Settings relating the fluid pressure
It is possible to set whether the absolute or the relative pressure is to be used:
• Select on or off.
• Press ENTER.
If on is selected, the absolute pressure p
a
will be displayed/input/transmitted.
If off is selected, the relative pressure p
g
will be displayed/input/transmitted.
p
g
= p
a
- 1.01 bar
The pressure and its unit of measurement will, e.g., be displayed in the program branch Parameter. It is followed by the
selected pressure, indicated in parentheses:
a – absolute pressure
g – relative pressure
Special Funct.\...\Dialogs/Menus\SHOW RELAIS STAT
Special Funct.\...\Dialogs/Menus\Pressure absolut
Notice!Notice!
The standard pressure in Special Funct.\SYSTEM settings\Gas-Measuring\Norm/Std press. is
entered as absolute value.
Fluid pressure
1.00 bar(a)