18 Settings
18.1 Dialogs and menus FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
158
18.1.6 Alarm state indication
Select on to display the alarm state during the measurement.
Press ENTER.
18.1.7 Units of measurement
It is possible to set the units of measurement for the length, temperature, pressure, density and kinematic viscosity and
sound speed.
Select a unit of measurement for all quantities.
Press ENTER after each selection.
18.1.8 Settings relating the fluid pressure
It is possible to set whether the absolute or the relative pressure is to be used:
Select on or off.
Press ENTER.
If on is selected, the absolute pressure p
a
will be displayed/input/transmitted.
If off is selected, the relative pressure p
g
will be displayed/input/transmitted.
p
g
= p
a
- 1.01 bar
The pressure and its unit of measurement will, e.g., be displayed in the program branch Parameter. It is followed by the
selected pressure, indicated in parentheses:
a – absolute pressure
g – relative pressure
Special Funct.\...\Dialogs/Menus\SHOW RELAIS STAT
Special Funct.\...\Dialogs/Menus\Pressure absolut
Notice!Notice!
The standard pressure in Special Funct.\SYSTEM settings\Gas-Measuring\Norm/Std press. is
entered as absolute value.
Fluid pressure
1.00 bar(a)