9 Measurement
9.3 Measurement settings FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
116
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing or further information is displayed.
Press ENTER.
Output function test
The function of the output can now be tested.
Connect an external measuring instrument to the terminals of the output.
•Select Yes to test the output. Select No to display the next menu item.
Press ENTER.
Enter a test value. It has to be within the output range.
Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
•Select Repeat to repeat the test, Finish to display the next menu item.
Press ENTER.
•Select Yes to test the assignment of the measured value to the output signal. Select No to display the next menu item.
Press ENTER.
Select in the scroll list Active or Passive.
Press ENTER.
If the external measuring instrument displays the value (min. output value for Passive, max. output for Active), the
output functions correctly.
•Select Repeat to repeat the test. Select Finish to display the next menu item.
Press ENTER.
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\...\Test signal
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Test mea. range
Options\Channel A\Outputs\...\Enter test value