9 Measurement
9.6 Display of measured values FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
126
Transducer distance
By pressing during the measurement, it is possible to scroll to the display of the transducer distance.
The recommended transducer distance will be displayed in brackets, the measured transducer distance will be displayed
afterwards. The recommended transducer distance might change during the measurement (e.g., due to temperature
fluctuations). A deviation from the recommended transducer distance will be compensated internally.
Transducer temperature
In the SuperUser and SuperUser ext. modes it is possible to display the transducer temperature during the
measurement.
By pressing during the measurement, it is possible to scroll to the transducer temperature display.
HPI physical quantities
The following physical quantities are additionally displayed in the HPI measuring mode.
Fig. 9.7: Display of the transducer distance
1 transducer distance
Notice!
Never change the transducer distance during the measurement.
Fig. 9.8: Display of the transducer temperature
Notice!
If the compliance of the specified transducer temperature has to be monitored, an event trigger can be set on the
temperature value.
Tab. 9.10: HPI physical quantities
display description
Df fluid density
Tf fluid temperature
Dfn standardized density
Fl fluid number
VCF volume correction factor
SLP slope
K API gravity