Annex
A Menu structure FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
216
Example 1
The diagnostic values are to be checked.
If the SCNR is < 20 dB, a snap has to be triggered.
Trigger:
Condition:
Action:
SCNR ˂ 20 dB
R1 with SCNR < 20 dB
trigger a snap
Options Special functions
Channel A Snap
Event trigger
Configuration
Select R1 On
Enable R1 Snap ringbuffer
Yes Yes
Source item Auto snap
→ Diagnostic values
No
Diagnostic values
→ SCNR Snap on R1
Function Yes
→ MIN (x<limit)
Type
→ Non-hold
Trigger value
→ 20.0 dB
Hysteresis
→ 1.0 dB
Glitch interval
→ 1.0 s
Failure delay
→ 1.0 s