EasyManua.ls Logo

Flexim FLUXUS H721 - Page 216

Flexim FLUXUS H721
229 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Annex
A Menu structure FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
216
Example 1
The diagnostic values are to be checked.
If the SCNR is < 20 dB, a snap has to be triggered.
Trigger:
Condition:
Action:
SCNR ˂ 20 dB
R1 with SCNR < 20 dB
trigger a snap
Options Special functions
Channel A Snap
Event trigger
Configuration
Select R1 On
Enable R1 Snap ringbuffer
Yes Yes
Source item Auto snap
Diagnostic values
No
Diagnostic values
SCNR Snap on R1
Function Yes
MIN (x<limit)
Type
Non-hold
Trigger value
20.0 dB
Hysteresis
1.0 dB
Glitch interval
1.0 s
Failure delay
1.0 s
1
2

Table of Contents

Related product manuals