6 Mounting
6.2 Transducers FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
42
• Select the installation instruction of the supplied clasp:
Band clamp clasp
• Insert the tension strap into the tension strap clamp (2).
• Place the tension strap around the pipe and insert it into the clasp (see Fig. 6.25 and Fig. 6.26).
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nut of tension strap clamp (2), but not too firmly in order not to damage the tension strap.
Quick release clasp
• Insert the tension strap into its clamp (2) and the metal spring.
• Place the tension strap around the pipe and insert it into the clasp (see Fig. 6.25 and Fig. 6.27).
• Position the metal spring opposite the tension strap clamp (2).
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nut of tension strap clamp (2), but not too firmly in order not to damage the tension strap.
Fig. 6.25: Rail on the pipe
1 – tension strap clamp
2 – tension strap clamp
3–metal spring
4 – nut
5–clasp
Fig. 6.26: Tension strap with band clamp
clasp
Fig. 6.27: Tension strap with quick release
clasp and metal spring
1 – clasp screw 1 – clasp screw
2–metal spring