EasyManua.ls Logo

Flexim FLUXUS H721 - Page 41

Flexim FLUXUS H721
229 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Mounting
FLUXUS H721 6.2 Transducers
41
UMFLUXUS_H721V1-5EN, 2022-05-15
Step 4: Fixation of the rail to the pipe
Place the tension strap clamp (2) in the rail (see Fig. 6.23). Observe the orientation of the tension strap clamp.
Slightly tighten the nut of the tension strap clamp (2).
Screw the rail to tension strap clamp (1) (see Fig. 6.24).
Tighten the nut of tension strap clamp (1), but not too firmly in order not to damage the tension strap.
Fig. 6.21: Tension strap with metal spring and clamp
1–metal spring
2 tension strap clamp
Fig. 6.22: Ratchet clasp with tension strap
1 sense of rotation
2 edge
3–lever
4 clasp screw with slot
Fig. 6.23: Rail with tension strap clamp
1 nut
2 tension strap clamp
Fig. 6.24: Rail, fixed on one side to the pipe
1 tension strap clamp
2 tension strap clamp
3 nut
12
1
2
4
3
4
1
2
213
1

Table of Contents