6 Mounting
6.3 Temperature probe FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
64
6.3.2.2 Installation with FLEXIM clasp
Cut the tension strap to length (pipe circumference + at least 120 mm).
Insert approx. 20 mm of the tension strap into the slot of the clasp.
Bend the end of the tension strap.
Position the temperature probe on the pipe (see Fig. 6.78).
Place the tension strap around the temperature probe and the pipe.
Push the tension strap through part (2) and (1) of the clasp.
Tighten the tension strap and engage it in the inner hook of the clasp.
Tighten the screw of the clasp.
Fig. 6.77: Clasp
Fig. 6.78: Temperature probe on the pipe
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
Debur sharp edges.
Wear the required personal protective equipment.
Observe the applicable rules.
Fig. 6.79: FLEXIM clasp
ab
(2)
(2)
(1)
(1)
(1)
(2)